WEBVTT

00:00:00.400 --> 00:00:06.124
Double-clicking the spectrum locates the nearby maximum and opens its attenuation.

00:00:06.124 --> 00:00:09.767
Check the marked position and extraction half-width.

00:00:09.767 --> 00:00:18.093
The same region is used throughout the ramp; the maximum is not chased in each spectrum.

00:00:18.093 --> 00:00:23.817
Here we use a project copy without changing the saved map.

00:00:25.817 --> 00:00:33.814
The D-only option fits monoexponential attenuation and reports measured diffusion, final diffusion and fit statistics.

00:00:33.814 --> 00:00:38.079
Check the internal reference and temperature normalization used.

00:00:38.079 --> 00:00:42.877
A high coefficient of determination does not demonstrate monodispersity.

00:00:42.877 --> 00:00:49.275
This fit alone does not produce a distribution or identify a compound.

00:00:51.275 --> 00:00:57.025
Switch to the linear view to inspect log normalized intensity against b.

00:00:57.025 --> 00:01:01.338
Review valid and excluded observations, the fit and residuals.

00:01:01.338 --> 00:01:05.650
Do not remove observations only to improve R squared.

00:01:05.650 --> 00:01:13.317
Any manual or automatic exclusion must retain a visible criterion and the observations no longer used.

00:01:15.317 --> 00:01:21.966
Calculate distribution inverts this region's attenuation using the selected method and grid.

00:01:21.966 --> 00:01:26.952
Review convergence, residuals, boundaries and moments before accepting it.

00:01:26.952 --> 00:01:33.601
Hydrodynamic radius requires documented viscosity and temperature; mass requires a compatible calibration.

00:01:33.601 --> 00:01:41.358
Adding a result is a separate action and is not performed in this video.