In internal references, review candidates and confirm the standard actually added.

00:00:06.936 --> 00:00:11.689
Check its region, attenuation, hydrodynamic radius and provenance.

00:00:11.689 --> 00:00:17.037
Detecting a signal does not automatically activate a reference.

00:00:17.037 --> 00:00:24.761
If you work without an internal standard, the result must retain that warning.

00:00:24.761 --> 00:00:30.108
This walkthrough uses demonstration settings, not a validated calibration.

00:00:32.108 --> 00:00:37.838
In molecular-weight analysis, review the inversion engine and mass model.

00:00:37.838 --> 00:00:44.712
Changing engine creates a different protocol; it does not rewrite historical results.

00:00:44.712 --> 00:00:51.015
Linear-polymer calibrations do not validate absolute masses of stars or brushes.

00:00:51.015 --> 00:00:57.317
Distinguish measured diffusion, corrected diffusion and apparent mass before comparing samples.

00:00:59.317 --> 00:01:06.853
Open advanced mode only when you need to inspect the grid, regularization and stopping criteria.

00:01:06.853 --> 00:01:13.384
Standard values support a reproducible workflow but do not guarantee convergence or identifiability.

00:01:13.384 --> 00:01:18.910
Do not relax a check merely to obtain an attractive figure.

00:01:18.910 --> 00:01:25.442
Retain the method, parameters, warnings and calculation status in the revision you accept.

00:01:27.442 --> 00:01:31.309
Review probe, calibration, solvent, viscosity and temperature.

00:01:31.309 --> 00:01:45.121
In the principal KK2 protocol, the TTMS reference factor and thermal factor are retained per block and applied together exactly once to the complete axis.

00:01:45.121 --> 00:01:51.198
Do not replace the experimental TBO calibration with a nominal gradient.

00:01:51.198 --> 00:01:57.275
This video shows where to inspect conditions without recalculating original data.